Application of the TPB Wavelength Shifter to the DEAP-3600 Spherical Acrylic Vessel Inner Surface

B. Broerman,M.G. Boulay,B. Cai,D. Cranshaw,K. Dering,S. Florian,R. Gagnon,P. Giampa,C. Gilmour,C. Hearns,J. Kezwer,M. Kuźniak,T. Pollmann,M. Ward
DOI: https://doi.org/10.1088/1748-0221/12/04/p04017
2017-04-18
Journal of Instrumentation
Abstract:DEAP-3600 uses liquid argon contained in a spherical acrylic vessel as a target medium to perform a sensitive spin-independent dark matter search. Argon scintillates in the vacuum ultraviolet spectrum, which requires wavelength shifting to convert the VUV photons to visible so they can be transmitted through the acrylic light guides and detected by the surrounding photomultiplier tubes. The wavelength shifter 1,1,4,4-tetraphenyl-1,3-butadiene was evaporatively deposited to the inner surface of the acrylic vessel under vacuum. Two evaporations were performed on the DEAP-3600 acrylic vessel with an estimated coating thickness of 3.00 ± 0.02 μm which is successfully wavelength shifting with liquid argon in the detector. Details on the wavelength shifter coating requirements, deposition source, testing, and final performance are presented.
instruments & instrumentation
What problem does this paper attempt to address?