Position Identification for Robust Myoelectric Control Against Electrode Shift

Jiayuan He,Xinjun Sheng,Xiangyang Zhu,Ning Jiang
DOI: https://doi.org/10.1109/tnsre.2020.3038374
IF: 4.9
2020-12-01
IEEE Transactions on Neural Systems and Rehabilitation Engineering
Abstract:The vulnerability to the electrode shift was one of the key barriers to the wide application of pattern recognition-based (PR-based) myoelectric control systems outside the controlled laboratory conditions. To overcome this challenge, a novel framework named position identification (PI) was proposed. In the PI framework, an anchor gesture performed by the user was first analyzed to identify the current electrode position from a pool of potential electrode shift positions. Next, the classifier calibrated by the data of the identified position would be selected for following myoelectric control tasks. The results of the amputee and able-bodied participants both demonstrated that the differential filter combined with majority voting improved the PI accuracy. With only one second contraction of the chosen anchor gesture (hand close), the subsequent PR-based myoelectric control performance was fully restored from eight different electrode shift scenarios, with 1 cm in either or both perpendicular and parallel directions. The classification accuracies with PI framework were not significant before and after the shift ( ${p}< $ 0.001). The advantage of restoring performance fully in just one second made it a practical solution to improve the robustness of PR-based myoelectric control systems in a wide range of real-world applications.
engineering, biomedical,rehabilitation