Image quality of opportunistic breast examinations in photon-counting computed tomography: A phantom study

S. Sawall,E. Baader,J. Wolf,J. Maier,H.-P. Schlemmer,S.O. Schönberg,I. Sechopoulos,M. Kachelrieß
DOI: https://doi.org/10.1016/j.ejmp.2024.103378
IF: 3.119
2024-05-28
Physica Medica
Abstract:Highlights • The resolution of PCCT systems might allow for opportunistic breast examinations. • Calcifications down to a size of 0.29 mm and fibers down to a size of 0.23 mm could be identified in PCCT and BCT. • Clinical PCCT could emerge as an alternative to dedicated breast. Purpose To compare the breast imaging performance of a clinical whole-body photon-counting CT (PCCT) to that of a dedicated breast CT (BCT) to determine the image quality of opportunistic breast examinations in clinical PCCT. Materials and methods To quantify image quality for breast cancer applications, acquisitions of a breast phantom including representations of calcifications, fibers, and masses were performed using a clinical PCCT and a dedicated BCT. When imaging with the PCCT, the phantom was also combined with a thorax phantom to simulate realistic patient positioning, while only the breast phantom was imaged in the BCT. Images in BCT were acquired at 7.0 mGy (CTDI 16 cm ) and using 2.6 mGy-25.0 mGy in the PCCT. Spatial resolution between the BCT and PCCT images was matched and data were reconstructed using the default methods of each system. The dose-normalized contrast-to-noise ratio (CNRD) of masses and the structural visibility of fibers and calcifications were evaluated as figures of merit for all reconstructions. Results CNRD between masses and background was 0.56 mGy -1⁄2 , on average with BCT and varied between 0.39 mGy -1⁄2 to 1.46 mGy -1⁄2 with PCCT over all dose levels, phantom configurations, and reconstruction algorithms. Calcifications down to a size of 0.29 mm and fibers down to a size of 0.23 mm could be reliably identified in the images of both systems. Conclusions Clinical PCCT provides an image quality superior to that obtained with BCT in terms of CNRD and allows for the identification of calcifications and fibers at comparable dose levels.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?