Contrast-Enhanced Ultrasound of the Pleural Cavity: A Method to Locate Pleural Catheters and Identify Fibrous Septa
Ke Bi,Bin Wang,Yi Zhang,Meng-Jun Shen,Hong-Wei Chen,Hui-Ming Zhu,Chun-Hong Tang,Yin Wang,Meng-jun Shen,Hong-wei Chen,Hui-ming Zhu,Chun-hong Tang
DOI: https://doi.org/10.1016/j.ultrasmedbio.2021.01.011
2021-05-01
Abstract:<p>We aimed to explore the value of contrast-enhanced ultrasound (CEUS) of the pleural cavity in locating catheters and identifying fibrous septa and to compare CEUS with multiple existing methods. We included 304 participants whose pleural effusion could not continue to be drained and compared the catheter-localization capabilities of empirical diagnosis, B-mode ultrasound with normal saline and CEUS, with computed tomography as the reference standard. CEUS performed the best (accuracy, 100%; sensitivity, 100%; specificity, 100%), followed by B-mode ultrasound with normal saline (accuracy, 77.78%; sensitivity, 62.5%; specificity, 100%), and finally empirical diagnosis (accuracy, 54.17%; sensitivity, 66.67%; specificity, 33.33%). The capabilities of CEUS and computed tomography to identify fibrous septa were evaluated, with B-mode ultrasound as the reference, and CEUS (accuracy, 100%; sensitivity, 100%; specificity, 100%) was superior to computed tomography (accuracy, 82.41%; sensitivity, 26.09%; specificity, 97.65%). Overall, CEUS can accurately locate catheters and identify fibrous septa, with performance superior to existing methods.</p>