A carbon nanotube x-ray source array designed for a new multisource cone beam computed tomography scanner

Boyuan Li,Christina R Inscoe,Shuang Xu,Timothy Capo,Donald A Tyndall,Yueh Z Lee,Jianping Lu,Otto Zhou
DOI: https://doi.org/10.1088/1361-6560/ad3323
IF: 3.5
2024-03-13
Physics in Medicine and Biology
Abstract:Cone beam computed tomography (CBCT) is known to suffer from strong scatter and cone beam artifacts. The purpose of this study is to develop and characterize a rapidly scanning carbon nanotube (CNT) field emission x-ray source array to enable a multisource CBCT (ms-CBCT) image acquisition scheme which has been demonstrated to overcome these limitations. A CNT x-ray source array with eight evenly spaced focal spots was designed and fabricated for a medium field of view ms-CBCT for maxillofacial imaging. An external multisource collimator was used to confine the radiation from each focal spot to a narrow cone angle. For ms-CBCT imaging, the array was placed in the axial direction and rapidly scanned while rotating continuously around the object with a flat panel detector. The x-ray beam profile, temporal and spatial resolutions, energy and dose rate were characterized and evaluated for maxillofacial imaging. The CNT x-ray source array achieved a consistent focal spot size of 1.10±0.04mm × 0.84±0.03mm and individual beam cone angle of 2.4°±0.08 after collimation. The x-ray beams were rapidly switched with a rising and damping times of 0.21ms and 0.19ms, respectively. Under the designed operating condition of 110kVp and 15mA, a dose rate of 8245 μGy/s was obtained at the detector surface with the inherent Al filtration and 2312 μGy/s with an additional 0.3mm Cu filter. There was negligible change of the x-ray dose rate over many operating cycles. A ms-CBCT scan of an adult head phantom was completed in 14.4s total exposure time for the imaging dose in the range of that of a clinical CBCT scanner. A spatially distributed CNT x-ray source array was designed and fabricated. It has enabled a new multisource CBCT to overcome some of the main inherent limitations of the conventional CBCT.
engineering, biomedical,radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?