Fabrication of single-crystal diamond neutron detector and its application in 14.1 MeV neutron detection in deuteriumtritium fusion experiment

Yi Yu,Ping Xu,Haiyang Zhou
DOI: https://doi.org/10.1088/2058-6272/acb48a
2023-01-18
Plasma Science and Technology
Abstract:A single-crystal diamond detector is fabricated to diagnose 14.1 MeV deuterium-tritium (D-T) fusion neutrons. The size of its diamond film is 4.5mm×4.5mm×500μm. This film is then sandwiched by a flat and a strip-pattern gold electrode. The dark current of this detector is experimentally measured to be lower than 0.1 nA under electric field of 30kV/cm. This diamond detector is used to measure the D-T fusion neutrons which has a flux of about 7.5×105/(s·cm2). The pronounced peak with a central energy of 8.28 MeV which characterizing 12C(n,α)9Be reaction in the neutron energy spectrum is experimentally diagnosed, and the energy resolution is better than 1.69%, which is the best result reported so far by using a diamond detector. A clear peak with a central energy of 6.52 MeV which characterizing 12C(n,n')3α reaction is also identified with an energy resolution of better than 7.67%.