Electrochemical HOCl Production Modeling for an Electrochemical Catheter

Dilara Ozdemir,Derek Fleming,Cristian Picioreanu,Robin Patel,Haluk Beyenal
DOI: https://doi.org/10.1149/1945-7111/ad8aee
IF: 3.9
2024-10-25
Journal of The Electrochemical Society
Abstract:Hypochlorous acid (HOCl) is a strong oxidizing agent that damages cells by interacting with lipids, nucleic acids, sulfur-containing amino acids, and membrane components. It is an endogenous substance produced by the immune system to protect mammals from pathogens. Previously, we developed an HOCl-generating electrochemical catheter (e-catheter) and demonstrated its ability to prevent central line-associated bloodstream infections. The e-catheter is an electrochemical system consisting of two parts - an e-hub and a tube. Working, counter, and reference electrodes are placed in the e-hub, which contains 0.9% NaCl as an electrolyte. Although a simple prototype of this device has shown activity against pathogens, it is helfup to understand the factors influencing associated electrochemical/chemical processes to optimize design and efficacy. A mathematical model with realistic catheter geometry could predict factors influencing HOCl generation and distribution in the catheter and could aid in optimizing these devices. Here, we developed what refer to as an Electrochemical Hypochlorous Acid Production (EHAP) model to predict factors influencing electrochemical generation and distribution of HOCl in e-catheters, including polarization time, diffusion of HOCl into the e-catheter, operational voltage, working electrode length, and surface area.
electrochemistry,materials science, coatings & films
What problem does this paper attempt to address?