A Portable Readout System for Biomarker Detection with Aptamer-Modified CMOS ISFET Array

Dmitriy Ryazantsev,Mark Shustinskiy,Andrey Sheshil,Alexey Titov,Vitaliy Grudtsov,Valerii Vechorko,Irakli Kitiashvili,Kirill Puchnin,Alexander Kuznetsov,Natalia Komarova
DOI: https://doi.org/10.3390/s24103008
IF: 3.9
2024-05-10
Sensors
Abstract:Biosensors based on ion-sensitive field effect transistors (ISFETs) combined with aptamers offer a promising and convenient solution for point-of-care testing applications due to the ability for fast and label-free detection of a wide range of biomarkers. Mobile and easy-to-use readout devices for the ISFET aptasensors would contribute to further development of the field. In this paper, the development of a portable PC-controlled device for detecting aptamer-target interactions using ISFETs is described. The device assembly allows selective modification of individual ISFETs with different oligonucleotides. Ta2O5-gated ISFET structures were optimized to minimize trapped charge and capacitive attenuation. Integrated CMOS readout circuits with linear transfer function were used to minimize the distortion of the original ISFET signal. An external analog signal digitizer with constant voltage and superimposed high-frequency sine wave reference voltage capabilities was designed to increase sensitivity when reading ISFET signals. The device performance was demonstrated with the aptamer-driven detection of troponin I in both reference voltage setting modes. The sine wave reference voltage measurement method reduced the level of drift over time and enabled a lowering of the minimum detectable analyte concentration. In this mode (constant voltage 2.4 V and 10 kHz 0.1Vp-p), the device allowed the detection of troponin I with a limit of detection of 3.27 ng/mL. Discrimination of acute myocardial infarction was demonstrated with the developed device. The ISFET device provides a platform for the multiplexed detection of different biomarkers in point-of-care testing.
engineering, electrical & electronic,instruments & instrumentation,chemistry, analytical
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to develop a portable read - out system for biomarker detection based on ion - sensitive field - effect transistors (ISFETs) and aptamer - modified complementary metal - oxide - semiconductor (CMOS) arrays. Specifically, the study aims to: 1. **Improve portability and ease of use**: Develop a portable, PC - controlled device for detecting the interaction between aptamers and target molecules. This device can be conveniently applied to point - of - care testing (POCT). 2. **Optimize the ISFET structure**: Optimize the ISFET structure of the Ta<sub>2</sub>O<sub>5</sub> gate to reduce trap charges and capacitance attenuation and improve the performance of the sensor. 3. **Integrate the read - out circuit**: Use an integrated CMOS read - out circuit with a linear transfer function to minimize the distortion of the original ISFET signal. 4. **Increase sensitivity**: Design an external analog - signal digitizer with the ability to have a constant voltage and superimpose a high - frequency sine - wave reference voltage to improve the sensitivity of reading ISFET signals. 5. **Verify device performance**: Verify the performance of the device by detecting cardiac troponin I. The study found that using the sine - wave reference - voltage measurement method can reduce drift and lower the minimum detectable analyte concentration. 6. **Multiplex detection**: Provide a platform to achieve multiplex detection of different biomarkers, especially suitable for POCT applications. Through these improvements, this study aims to promote the practical application of ISFET biosensors in the early diagnosis of cardiovascular diseases and other medical fields.