An international phantom study of inter-site variability in Technetium-99m image quantification: analyses from the TARGET radioembolization study

Grace Keane,Rob van Rooij,Marnix Lam,S. Cheenu Kappadath,Bilal Kovan,Stephanie Leon,Matthew Dreher,Kirk Fowers,Hugo de Jong
DOI: https://doi.org/10.1186/s40658-024-00647-x
2024-05-30
EJNMMI Physics
Abstract:Personalised multi-compartment dosimetry based on [ 99m Tc]Tc-MAA is a valuable tool for planning 90 Y radioembolization treatments. The establishment and effective application of dose–effect relationships in yttrium-90 ( 90 Y) radioembolization requires [ 99m Tc]Tc-MAA SPECT quantification ideally independent of clinical site. The purpose of this multi-centre phantom study was to evaluate inter-site variability of [ 99m Tc]Tc-MAA imaging and evaluate a standardised imaging protocol. Data was obtained from the TARGET study, an international, retrospective multi-centre study including 14 sites across 8 countries. The impact of imaging related factors was estimated using a NEMA IQ phantom (representing the liver), and a uniformly filled cylindrical phantom (representing the lungs). Imaging was performed using site-specific protocols and a standardized protocol. In addition, the impact of implementing key image corrections (scatter and attenuation correction) in the site-specific protocols was investigated. Inter-site dosimetry accuracy was evaluated by comparing computed Lung Shunt Fraction (LSF) measured using planar imaging of the cylindrical and NEMA phantom, and contrast recovery coefficient (CRC) measured using SPECT imaging of the NEMA IQ phantom.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?