Initial Experience with Duodenoscopes with Single-Use End Caps in Pediatric Endoscopic Retrograde Cholangiopancreatography: Infection Prevention Comes at a Cost.

Monique T. Barakat,Andrew Liman,Roberto Gugig
DOI: https://doi.org/10.1016/j.gie.2024.03.009
IF: 10.396
2024-03-10
Gastrointestinal Endoscopy
Abstract:Background and Aims Duodenoscopes with single-use end caps were introduced to minimize infection risk, but are unstudied in pediatrics. Methods We collected clinical data and endoscopists' evaluations of duodenoscopes with single-use end caps versus reusable duodenoscopes over 18 months. Results A total of 106 ERCPs were performed for patients aged 1-18 (mean 14.2) years. Forty-six involved single-use end caps, with 9 requiring crossover to reusable duodenoscopes. ERCPs involving single-use end caps resulted in more instances of mucosal trauma (10 vs 0, p<0.05) and post-ERCP pancreatitis (4 vs 1, p<0.05), and accounted for 8 of 9 ERCPs requiring advanced cannulation techniques. No post-ERCP infections occurred. Reported challenges included single-use end cap stiffness and difficulty with their alignment for cannulation. Conclusions We report difficulty with advancement, greater reliance on advanced cannulation techniques, and higher rates of PEP when using duodenoscopes with single-use end caps in pediatric ERCP. This area warrants further study.
gastroenterology & hepatology
What problem does this paper attempt to address?