Signal-correction errors in the EasyOne Pro LAB multiple-breath washout device significantly impact outcomes in children and adults

Marc-Alexander Oestreich,Florian Wyler,Christian Buess,Blaise Etter,Kathryn A Ramsey,Philipp Latzin,Kathryn A. Ramsey
DOI: https://doi.org/10.1152/japplphysiol.00096.2022
2024-01-25
Journal of Applied Physiology
Abstract:Background: Multiple-breath washout (MBW) is an established technique to assess functional residual capacity (FRC) and ventilation inhomogeneity in the lung. Indirect calculation of nitrogen concentration requires accurate measurement of gas concentrations. Aim: To investigate the accuracy of the CO 2 concentration and molar mass (MM) values used for the indirect calculation of nitrogen concentration in a commercial MBW device (EasyOne Pro LAB (EOPL), ndd Medizintechnik AG, Switzerland) and its impact on outcomes. Methods: We used high-precision gas mixtures to evaluate CO 2 and MM sensor output in-vivo and in-vitro. We developed updated algorithms to correct for observed errors and assessed the impact on MBW outcomes and FRC measurement accuracy compared to body plethysmography. Results: The respiratory exchange ratio (RER)-based adjustment of the measured CO 2 signal used in the EOPL led to an overestimated CO 2 signal (range -0.1; 1.0%). In addition, an uncorrected dependence on humidity was identified. These combined effects resulted in an overestimation of expired nitrogen concentrations (range -0.7; 2.6%), and consequently MBW outcomes. Corrected algorithms reduced the mean (SD) cumulative expired volume by 15.8% (9.7%), FRC by 6.6% (3.0%), and lung clearance index by 9.9% (7.6%). Differences in FRC between the EOPL and body plethysmography further increased. Conclusion: Inadequate signal correction causes RER- and humidity-dependent expired nitrogen concentration errors and overestimation of test outcomes. Updated algorithms reduce average signal error, however, RER values far from the population average still cause measurement errors. Despite improved signal accuracy, the updated algorithm increased the difference in FRC between the EOPL and body plethysmography.
physiology,sport sciences
What problem does this paper attempt to address?