Fat Suppression in Distal Extremity 3-T MRI Using Spectral Heterogeneity Adaptive Radiofrequency Pulses

Iman Khodarahmi,William R Walter,Mary Bruno,Inge M Brinkmann,Mahesh B Keerthivasan,Venkata V Chebrolu,Jan Fritz
DOI: https://doi.org/10.1148/radiol.231184
IF: 19.7
2024-09-25
Radiology
Abstract:Background Conventional chemical shift selective (CHESS) fat suppression may fail in distal extremity MRI due to sensitivity to field inhomogeneities. Purpose To develop a patient-specific fat-suppression method for distal extremity 3-T MRI by exploiting the spectral heterogeneity adaptive radiofrequency pulse (SHARP) technique and to compare it to fat suppression with CHESS. Materials and Methods SHARP uses the routinely acquired frequency spectrum at MRI calibration to adapt the frequency...