Effects on mechanical power of different devices used for inhaled sedation in a bench model of protective ventilation in ICU

Pierre-Louis Pellet,Neven Stevic,Florian Degivry,Bruno Louis,Laurent Argaud,Claude Guérin,Martin Cour
DOI: https://doi.org/10.1186/s13613-024-01245-x
IF: 10.318
2024-01-31
Annals of Intensive Care
Abstract:Inhaled sedation during invasive mechanical ventilation in patients with acute respiratory distress syndrome (ARDS) has received increasing attention. However, inhaled sedation devices increase dead-space ventilation and an undesirable effect is the increase in minute ventilation needed to maintain CO 2 removal. A consequence of raising minute ventilation is an increase in mechanical power (MP) that can promote lung injury. However, the effect of inhaled sedation devices on MP remains unknown.