Design and fabrication of a robotic mechanism for remote steering and positioning of interventional devices

Govindarajan Srimathveeravalli,Thenkurussi Kesavadas,Xinyan Li
DOI: https://doi.org/10.1002/rcs.301
2010-01-01
Abstract:BACKGROUND: Recently, robotic systems have been introduced as a method for assisting endovascular interventional procedures. A few commercial and research solutions are available. In a survey it was found that none of the current systems satisfied all basic design requirements set forth for a good robot-assisted therapy platform.METHODS: A human motion analysis study was performed to identify design specifications for safe motion and force limits for endovascular surgery. Based on design requirements from surveyed systems and motion analysis, a new, teleoperated, haptically enabled system called System for Endovascular Teleoperated Access (SETA) was constructed. SETA is capable of simultaneously manipulating any guidewire and catheter in the range of 0.014-0.13 inches.RESULTS: SETA's slave was evaluated for precision positioning, using in vitro vascular phantoms. It was also evaluated by a cohort of neurovascular surgeons and fellows (n = 8), using the VIST vascular simulator. A qualitative survey of the participants and a quantitative analysis of metrics procedure time and contrast used found that SETA was equivalent to manual intervention on this platform.CONCLUSIONS: SETA provided good performance in the in vitro studies, and will soon be evaluated in a series of in vivo animal model studies.
surgery
What problem does this paper attempt to address?