High precision weighing and uncertainty assessment in reagent preparation

Yang Zheng,Xujun Yuan,Rongfu Zhang,Qi Chen
DOI: https://doi.org/10.54254/2753-8818/33/20240913
2024-03-08
Abstract:The volume of the reagent in one tube was difficult to be measured in reagent preparation, as the volume of each portion was extremely small. Formerly, the measurement errors of direct weighing method were large. Therefore, the method of twice minusing method was used to weighing in order to obtain the reagent preparation result more accurately. The sources of uncertainty of this method were analyzed and the uncertainty was evaluated, the uncertainty of twice minusing method and direct weighing method was compared. The results showed that the expanded uncertainty of the reagents in one of the eight-connected tubes was 2.26610-4 g (k = 2) when using twice minusing method, and the expanded uncertainty was 1.70410-3 g (k = 2) when using direct weighing method, so the results of twice minusing method were more reliable.
What problem does this paper attempt to address?