Impact of Structural Compliance of a Six Degree of Freedom Joint Simulator on Virtual Ligament Force Calculation in Total Knee Endoprosthesis Testing

Eric Kleist,Paul Henke,Leo Ruehrmund,Maeruan Kebbach,Rainer Bader,Christoph Woernle
DOI: https://doi.org/10.3390/life14040531
2024-04-22
Life
Abstract:The AMTI VIVOTM six degree of freedom joint simulator allows reproducible preclinical testing of joint endoprostheses under specific kinematic and loading conditions. When testing total knee endoprosthesis, the articulating femoral and tibial components are each mounted on an actuator with two and four degrees of freedom, respectively. To approximate realistic physiological conditions with respect to soft tissues, the joint simulator features an integrated virtual ligament model that calculates the restoring forces of the ligament apparatus to be applied by the actuators. During joint motion, the locations of the ligament insertion points are calculated depending on both actuators' coordinates. In the present study, we demonstrate that unintended elastic deformations of the actuators due to the specifically high contact forces in the artificial knee joint have a considerable impact on the calculated ligament forces. This study aims to investigate the effect of this structural compliance on experimental results. While the built-in algorithm for calculating the ligament forces cannot be altered by the user, a reduction of the ligament force deviations due to the elastic deformations could be achieved by preloading the articulating implant components in the reference configuration. As a proof of concept, a knee flexion motion with varying ligament conditions was simulated on the VIVO simulator and compared to data derived from a musculoskeletal multibody model of a total knee endoprosthesis.
biology
What problem does this paper attempt to address?
The paper aims to address the issue of inaccurate virtual ligament force calculations in the six-degree-of-freedom joint simulator (AMTI VIVO™) during total knee prosthesis testing due to structural compliance. Specifically: 1. **Research Background**: Testing of total knee prostheses requires a highly precise simulation environment. The VIVO simulator is an advanced six-degree-of-freedom joint simulator capable of preclinical testing under physiological conditions. 2. **Problem Description**: In the presence of high contact forces in artificial knee joints, the actuators of the simulator undergo unexpected elastic deformation, which significantly affects the ligament forces calculated by the virtual ligament model. 3. **Research Objective**: This paper is divided into 2 parts. The first part demonstrates the impact of such structural compliance on virtual ligament forces. The second part proposes a method to reduce these effects by preloading the implant components to minimize ligament force deviations caused by elastic deformation. 4. **Experimental Validation**: The effectiveness of the method was validated through knee flexion experiments on the VIVO simulator and by comparing the data with a multi-body system model. Additionally, a clinically relevant loading case—virtual posterior cruciate ligament (PCL) resection—was considered to further analyze the impact of the joint simulator's structural compliance on experimental results.