Clinical Performance of Immunonephelometric Assay and Chemiluminescent Immunoassay for Detection of IgG Subclasses in Chinese

Yan Qin,Yuhan Jia,Congcong Liang,Rui Fu,Zhaojun Liang,Yanlin Wang,Min Feng,Chong Gao,Jing Luo
DOI: https://doi.org/10.1002/jcla.25033
IF: 3.124
2024-04-04
Journal of Clinical Laboratory Analysis
Abstract:A total of 40 individuals with IgG4‐RD, 40 with primary Sjogren's syndrome (pSS), and 40 healthy controls (HCs) were recruited. Immunoglobulin G (IgG) subclasses (IgG1, IgG2, IgG3, and IgG4) were parallelly detected by the Siemens immunonephelometric assay and the chemiluminescent immunoassay (CLIA). The levels of the total serum IgG were measured using the Beckman Coulter IMMAGE 800. The correlation analysis of two methods were carried out, and the diagnostic value was analyzed by the receiver operating characteristic (ROC) curve. Quantitative results from the CLIA correlated well with the Siemens immunonephelometric assay. The CLIA and the Siemens assay performed well in detecting IgG subclasses, especially IgG4 and IgG1 that can accurately identify patients with IgG4‐RD or pSS in clinical practice. Background Detection of IgG subclasses (IgGSc) is vital for the diagnosis and management of disease, especially IgG4‐related diseases (IgG4‐RD). This study aimed to evaluate the performances of the chemiluminescent immunoassay (CLIA) for detecting IgGSc and diagnosing IgG4‐RD by IgGSc. Methods A total of 40 individuals with IgG4‐RD, 40 with primary Sjogren's syndrome (pSS), and 40 healthy controls (HCs) were enrolled. Serum samples were collected for the simultaneous detection of IgG1, IgG2, IgG3, and IgG4 by the Siemens immunonephelometric assay and the CLIA. The correlation analysis was performed, and diagnostic value was analyzed by the receiver operating characteristic (ROC) curve. Results Patients with IgG4‐RD had higher IgG4 (p
medical laboratory technology
What problem does this paper attempt to address?