Fluorescence Spectra of Prototropic Forms of Fluorescein and Some Derivatives and Their Potential Use for Calibration-Free pH Sensing

Bernard Gauthier-Manuel,Chafia Benmouhoub,Bruno Wacogne
DOI: https://doi.org/10.3390/s24051705
IF: 3.9
2024-03-07
Sensors
Abstract:Fluorescence pH sensing has proven to be efficient but with the drawback that molecules photobleach, requiring frequent calibrations. Double-emission peak molecules allow ratiometric measurements and theoretically avoid calibration. However, they are often expensive and fragile and usually have very low quantum yields. Single emission peaks such as fluorescein and derivatives are inexpensive and have very high quantum yields. Because they are single emission peaks, the pH is assumed to be derived from the ratio of emitted intensities at measured pH and at high pH values, i.e., they require frequent calibration. However, the shape of their single emitted peak evolves slightly with pH. In this paper, we first demonstrate a simple method to calculate the emission spectrum shape of each prototropic form of fluorescein (and derivatives) as well as the values of the pKas. A complete model of the evolution of the emission spectrum shape with pH is then constructed. Second, we evaluate the potential of these molecules for pH sensing by fitting the experimental spectra with the complete emission model. The method is applied to fluorescein, FITC and FAM. Depending on the molecule, pH can be measured from pH 1.9 to pH 7.3 with standard deviations between 0.06 and 0.08 pH units. Estimating pH and pKas from shape instead of intensity allows calibration-free measurements even with single-emission peak molecules.
engineering, electrical & electronic,chemistry, analytical,instruments & instrumentation
What problem does this paper attempt to address?
The problem that this paper attempts to solve is the common calibration difficulties in fluorescent pH sensing. Specifically, in the process of using traditional fluorescent pH sensors, the concentration will decrease due to molecular photobleaching and other reasons, and frequent calibration is required. Although dual - emission - peak molecules allow ratio measurement and theoretically avoid calibration, these molecules are usually expensive and fragile, and also have a very low quantum yield. In contrast, single - emission - peak molecules such as fluorescein and its derivatives have the advantages of low cost and high quantum yield. However, since they have only one emission peak, the pH value is usually determined by measuring the ratio of the emission intensity at an unknown pH to that at a high pH, which also requires frequent calibration. The paper proposes a new method to achieve calibration - free pH measurement by analyzing the changes in the shape of a single emission peak of fluorescein and its derivatives at different pH values, rather than relying on the intensity ratio. This method is based on the calculation of the fluorescence emission spectral shape of each proton - isomeric form and the pKa value. By constructing a complete model to describe the law of the change of the emission spectral shape with pH and applying it to the fitting of experimental data, the paper evaluates the potential of these molecules as calibration - free pH sensors. In short, the main objective of the paper is to develop a new method for pH measurement using the changes in the emission spectral shape of fluorescein and its derivatives to overcome the calibration requirements in existing technologies.