The role of counter-torque holder in the tightening of pedicle screw-rod constructs: a biomechanical study in a porcine model.

Ming-Kai Hsieh,Po-Yi Liu,Yun-Da Li,Chi-Yun Wang,Chih-Chung Hu,Ching-Lung Tai,Po-Liang Lai
DOI: https://doi.org/10.1016/j.spinee.2022.08.021
2022-09-27
Abstract:Background context Pedicle screw-rod assembly procedures following pedicle screw insertion include contouring and placing rods into screw tulips, introducing set screws into the tulip along the screw thread, applying a counter-torque holder and tightening all the set screws clockwise. Even if an appropriate pedicle screw is implanted, screw dislodgement after tightening of the tulip and set screw is not uncommon. Pedicle wall violation resulting from excessive rotational force due to inadequate use of a counter-torque holder might be the reason. However, the strain change in the pedicle during tulip-set screw tightening and the role of counter-torque have never been investigated. Purpose This study determined differences in the strain change in the outer and inner pedicle walls during tulip-set screw tightening; additionally, the influence of counter-torque on pedicle wall violation was elucidated. Study design A controlled biomechanical study; the strain values of outer and inner pedicle walls in cadaveric porcine L4-L5 vertebrae during tulip-set screw tightening with or without a counter-torque holder were measured. Methods Twelve L4-L5 fresh-frozen porcine lumbar vertebrae were implanted with screw-rod constructs; the set screw was randomly locked into the tulip in the right L5, right L4, left L5 and left L4 testing groups. The maximal values from eight strain gauges (P-R-O: outer cortex of right pedicle in proximal vertebra; P-R-I: inner cortex of right pedicle in proximal vertebra; D-R-O: outer cortex of right pedicle in distal vertebra; D-R-I: inner cortex of right pedicle in distal vertebra; P-L-O: outer cortex of left pedicle in proximal vertebra; P-L-I: inner cortex of left pedicle in proximal vertebra; D-L-O: outer cortex of left pedicle in distal vertebra; D-L-I: outer cortex of left pedicle in proximal vertebra) for each specimen during tightening to 12 Nm were measured. Results The maximal strain values of the ipsilateral strain gauges in all testing groups were almost significantly higher when a counter-torque holder was not used than when one was used. The strain values in the adjacent pedicle of specimens without a counter-torque holder were significantly increased: P-R-O and P-R-I in the right L5 group; D-R-I in the right L4 group; P-L-I and P-L-O in the left L5 group; D-L-O and D-L-I in the left L4 group. Conclusions The constraint effect of counter-torque during tulip-set screw tightening is necessary. Clockwise rotational force with a fragile lateral pedicle wall suggests that caution is required when using a counter-torque holder to tighten the right L5 and left L4 constructs. Clinical significance A counter-torque holder is important during tulip-set screw tightening; improper use may lead to adjacent pedicle wall violation, sequentially resulting in pedicle screw loosening.
clinical neurology,orthopedics
What problem does this paper attempt to address?