Analytical evaluation of the automated interleukin-6 assay on the Roche cobas e602 analyzer

Anastasia Gant Kanegusuku,Timothy Carll,Kiang-Teck J Yeo
DOI: https://doi.org/10.1093/labmed/lmae064
IF: 1.865
2024-08-10
Laboratory Medicine
Abstract:Abstract Background Interleukin-6 (IL-6) is a proinflammatory cytokine that is associated with many inflammatory diseases. This validation study evaluates the automated Roche Elecsys IL-6 electrochemiluminescent immunoassay that has been granted emergency use authorization by the US Food and Drug Administration. Methods The Elecsys IL-6 assay was evaluated for precision, linearity, interference (by hemoglobin, bilirubin, triglycerides, and biotin) and clinical performance was compared to the V-PLEX Human IL-6 immunoassay (Meso Scale Discovery), performed by a reference laboratory. Results The Elecsys IL-6 assay is precise (intra-assay <3% coefficient of variation [CV], interassay <5% CV), exhibits an analytical measurable range of 1.5-4790 pg/mL, and is tolerant of significant interferences (H < 2522, I <62, L<2101, biotin <50 ng/mL). Comparison with the V-PLEX assay revealed a 2.95 slope bias in patient samples evaluated for IL-6 concentration (n = 43, range = 1.5-1891 pg/mL, y = 2.95x – 32.7, r2 = 0.84). Bland-Altman analysis revealed an absolute mean bias of 152 pg/mL (SD = 254 pg/mL), or a mean percentage difference of 73%. Conclusion The Roche IL-6 assay showed good analytical performance. The large systematic bias compared with another reference method precludes using multiple methods to monitor IL-6 response. The random-access nature of an automated IL-6 assay on the Roche platform makes the test available on demand.
medical laboratory technology
What problem does this paper attempt to address?