Evaluation of the humidification performances of new generation of heated wire humidifiers

François Lellouche,Emilie Rousseau,Pierre-Alexandre Bouchard
DOI: https://doi.org/10.4187/respcare.11352
2023-09-21
Respiratory Care
Abstract:BACKGROUND: At high ambient temperatures in ICU rooms, the humidification performances of heated wire humidifiers (HH) are significantly reduced, with delivered gas humidity well below 30 mgH2O/L leading to an increased risk of endo-tracheal occlusions, sub-occlusions or mucociliary dysfunction.The objective of the study was to evaluate the humidity delivered at the Y-piece with new generation HH with advanced algorithm (FP950 and VHB20) while varying ambient temperatures.METHODS: We measured on bench the hygrometry of inspiratory gases delivered by new generation of HH (i) FP950 (Fisher&Paykel Healthcare) (ii) VHB20 (Vincent Medical) and previous generation of HH (iii) MR850 with usual settings (37 at the chamber/40 at the Y piece) (iv) MR 850 with no temperature gradient (40/40), and (v) MR850 with automatic compensation algorithm activated. Hygrometry was measured with the psychrometric method after one hour of stability while varying the room temperature from 20 to 30°C.RESULTS: Two hundred and ninety-four hygrometric bench measurements were performed at steady state for the different tested conditions. With the new HH (FP950 and VHB20), gas humidity delivered remained above 30 mgH2O/L in all tested conditions, even at high ambient temperatures (> 25°C). With previous generation of HH (MR850), at high ambient temperature, humidity delivered was adequate in only 26% (11/42) of the measurements when usual settings were used (37/40) and 30% (11/37) with automatic compensation. When no temperature gradient was set (40/40), humidity delivered was above 30 mgH2O/L in 91% (30/33) of the measurements at high ambient temperature. With ambient temperature below 25°C, almost all devices and settings provided adequate humidity.CONCLUSION: The new FP950 and VHB20 heated wire heated humidifiers using advanced algorithms have demonstrated stable performance while varying the ambient temperature by 20-30°C, better than the previous generation of heated humidifiers when ambient temperatures were high.
respiratory system,critical care medicine
What problem does this paper attempt to address?