Prospective inter- and intra-tracer repeatability analysis of radiomics features in [68Ga]Ga-PSMA-11 and [18F]F-PSMA-1007 PET scans in metastatic prostate cancer

Jake Kendrick,Roslyn J. Francis,Ghulam Mubashar Hassan,Pejman Rowshanfarzad,Jeremy S.L. Ong,Robert Jeraj,Nathaniel Barry,Tammy Hagan,Martin A. Ebert,Roslyn Francis,Martin A Ebert
DOI: https://doi.org/10.1259/bjr.20221178
2023-09-27
British Journal of Radiology
Abstract:Objective: This study aimed to quantify both the intra- and inter tracer repeatability of lesion-level radiomics features in [ 68 Ga]Ga-PSMA-11 and [ 18 F]F-PSMA-1007 PET scans. Methods: Eighteen patients with metastatic prostate cancer (mPCa) were prospectively recruited for the study and randomised to one of three test-retest groups: (i) intra tracer [ 68 Ga]Ga-PSMA-11 PET, (ii) intra tracer [ 18 F]F-PSMA-1007 PET or (iii) inter tracer between [ 68 Ga]Ga-PSMA-11 and [ 18 F]F-PSMA-1007 PET. Four conventional PET metrics (SUV max , SUV mean , SUV total and volume) and 107 radiomics features were extracted from 75 lesions and assessed using the repeatability coefficient (RC) and the ICC. Radiomic feature repeatability was also quantified after the application of 16 filters to the PET image. Results: Test-retest scans were taken a median of 5 days apart (range: 2–7 days). SUV mean demonstrated the lowest RC limits of the conventional features, with RCs of 7.9%, 14.2 and 24.7% for the [ 68 Ga]Ga-PSMA-11 PET, [ 18 F]F-PSMA-1007 PET, and inter tracer groups, respectively. 69%, 66 and 9% of all radiomics features had good or excellent ICC values (ICC ≥0.75) for the same groups. Feature repeatability therefore diminished considerably for the inter tracer group relative to intra tracer groups. Conclusion: In this study, robust biomarkers for each tracer group that can be used in subsequent clinical studies were identified. Overall, the repeatability of conventional and radiomic features were found to be substantially lower for the inter tracer group relative to both intra tracer groups, suggesting that assessing patient response quantitatively should be done using the same radiotracer where possible. Advances in knowledge: Inter tracer biomarker repeatability limits are significantly larger than intra tracer limits.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?