Device profile of the Evoke physiologic closed-loop spinal cord stimulation system for the treatment of chronic intractable pain: overview of its safety and efficacy

Lawrence Poree,Allison Foster,Peter S. Staats
DOI: https://doi.org/10.1080/17434440.2023.2255520
IF: 3.439
2023-09-15
Expert Review of Medical Devices
Abstract:The Evoke® spinal cord stimulation (SCS) device enables the closed-loop feedback of dynamically measured evoked compound action potentials (ECAPs) to adjust stimulation amplitude for every stimulation pulse to maintain the stimulation output level near a targeted ECAP amplitude. No other commercially available SCS device presently uses physiologic feedback from the spinal cord to adjust stimulation. Clinicians should be familiar with the differences in devices and with the latest technologies to provide optimized patient care. In this device profile, the Evoke system is described and the system capabilities are differentiated from other available SCS devices. A systematic review was conducted based on best practice guidance to identify all available evidence on the safety and efficacy of the Evoke SCS system. The Evoke SCS system offers unique capabilities as a means to optimize therapy delivery tailored to each individual patient. Data through 24-months follow-up show statistically significant, clinically meaningful, ample, consistent, and strong evidence of the safety and efficacy of the Evoke system for the treatment of chronic intractable pain.
engineering, biomedical
What problem does this paper attempt to address?