Low-dose dynamic cerebral perfusion CT reconstruction based on voxel-level TAC correction (VTC)

Zixiang Chen,Na Zhang,Qiyang Zhang,Kaiyi Liang,Zhenxing Huang,Guotao Quan,Xiang Li,Dong Liang,Zhanli Hu
DOI: https://doi.org/10.1016/j.bspc.2023.105225
IF: 5.1
2023-07-02
Biomedical Signal Processing and Control
Abstract:Dynamic cerebral perfusion computed tomography (DCP-CT) is an advanced imaging technique that helps in the clinical diagnosis of cerebrovascular diseases (CVDs). However, radiation dose deposition during repeated CT scans seriously limits its clinical application. Effective DCP-CT imaging method based on low-dose protocol is needed. A regularized least-squares method with high interpretability based on voxel-level time-attenuation curve (TAC) correction (VTC) is proposed in this study for low-dose DCP-CT image reconstruction. The theory of third-order Hermite interpolation (THI) is applied to voxel-level TAC correction during dynamic images reconstruction. The peak signal-to-noise ratio (PSNR) and structural similarity (SSIM) are used to quantitatively evaluate the proposed method in terms of imaging accuracy and noise reduction, while hemodynamic maps, including cerebral blood flow (CBF), cerebral blood volume (CBV) and mean transition time (MTT), are calculated to validate its ability to restore hemodynamic parameters. It is proven that the proposed VTC method for low-dose DCP-CT imaging has better performance than several state-of-the-art dynamic CT imaging methods. It can be expected that the VTC method can play an important role in the promotion of the clinical application of DCP-CT for the diagnosis of CVD.
engineering, biomedical
What problem does this paper attempt to address?