Roderica R. G. Ng,Suneel R. Desai,Felicia S. W. Chu,Ming Ann Sim,Sheryl W. L. Chee,Jerry Y. H. Fuh,Lian-Kah Ti,Sophia T. H. Chew
Abstract:Oxygen delivery (DO 2 ) and its monitoring are highlighted to aid postoperative goal directed therapy (GDT) to improve perioperative outcomes such as acute kidney injury (AKI) after high-risk cardiac surgeries associated with multiple morbidities and mortality. However, DO 2 monitoring is neither routine nor done postoperatively, and current methods are invasive and only produce intermittent DO 2 trends. Hence, we proposed a novel algorithm that simultaneously integrates cardiac output (CO), hemoglobin (Hb) and oxygen saturation (SpO 2 ) from the Edwards Life Sciences ClearSight System® and Masimo SET Pulse CO-Oximetry® to produce a continuous, real-time DO 2 trend.
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to develop a new type of algorithm to achieve continuous non - invasive oxygen delivery (DO2) monitoring for patients after cardiac surgery. Specifically, the research aims to generate real - time DO2 trend graphs by simultaneously integrating the data provided by Edwards Life Sciences ClearSight System® and Masimo SET Pulse CO - Oximetry®, in order to improve the peri - operative management of cardiac surgery patients.
### Research Background
- **Complications of High - Risk Cardiac Surgery**: Approximately 2 million open - heart surgeries are performed globally each year, and these high - risk cardiac surgeries are associated with multiple complications and a relatively high mortality rate.
- **Importance of DO2 Optimization**: Optimizing oxygen delivery (DO2) has been proven to improve the peri - operative outcomes of cardiac surgery, especially in reducing acute kidney injury (AKI).
- **Limitations of Current DO2 Monitoring**: Current DO2 monitoring methods are neither routine nor performed post - operatively, and the existing methods are invasive and can only provide intermittent DO2 trends.
### Research Objectives
- **Develop a New Algorithm**: Propose a new algorithm to generate continuous, real - time DO2 trend graphs by simultaneously integrating data on cardiac output (CO), hemoglobin (Hb), and oxygen saturation (SpO2).
- **Verify Feasibility**: Evaluate the feasibility and preliminary effects of continuous non - invasive DO2 monitoring with this algorithm in cardiac surgery patients.
- **Personalized DO2 Thresholds**: Lay the foundation for future research, aiming to identify personalized critical DO2 thresholds and use them as part of goal - directed therapy (GDT) to optimize DO2 and improve the peri - operative outcomes of cardiac surgery.
### Main Contributions
- **Non - Invasive and Continuous**: This algorithm achieves non - invasive and continuous DO2 monitoring, solving the problems of invasiveness and intermittence in traditional methods.
- **Real - Time Monitoring**: Display the DO2 trend in real - time through a graphical user interface (GUI), enabling doctors to detect changes in patients' parameters in a timely manner.
- **Personalized Medicine**: Each patient can be their own control, with real - time monitoring of their DO2 trend, which is helpful for personalized medical decision - making.
### Research Methods
- **System Construction**: Build a real - time DO2 monitoring system through four components: machine interface, data extraction, data synchronization and integration, and real - time data display (GUI).
- **Data Processing**: Use Python code to synchronize and integrate the data from ClearSight System® and Masimo SET Pulse CO - Oximetry®, and calculate DO2.
- **Clinical Application**: Test this algorithm in 61 cardiac surgery patients, and record the DO2 trend and the occurrence of AKI.
### Results
- **Successful Algorithm Implementation**: The algorithm was successfully implemented in 57 patients, generating complete DO2 trend graphs.
- **Relationship between DO2 and AKI**: The overall DO2 values in patients with postoperative AKI were lower, but did not reach statistical significance.
- **SpHb Reading Verification**: Verify the accuracy of the SpHb readings of Masimo SET Pulse CO - Oximetry® by comparing them with laboratory Hb values.
### Discussion
- **Technical Advantages**: This algorithm is non - invasive, continuous, and can display DO2 trends in real - time, which is helpful for early detection of changes in patients' parameters.
- **Future Directions**: Further optimize the algorithm, incorporate user input to provide real - time exponential DO2 values, and develop an intelligent alarm system to improve the practicality and effectiveness of clinical applications.
In conclusion, this study demonstrates the feasibility and preliminary effects of achieving continuous non - invasive DO2 monitoring in cardiac surgery patients, providing new tools and technical support for future personalized medicine and goal - directed therapy.