NeuralDAO: Incorporating neural network generated dose into direct aperture optimization for end‐to‐end IMRT planning

Cong Liu,Xinye Ni,Xiance Jin,Wen Si
DOI: https://doi.org/10.1002/mp.15155
IF: 4.506
2021-08-25
Medical Physics
Abstract:Purpose: The current practice in Intensity Modulated Radiation Therapy (IMRT) planning almost always includes different dose calculation strategies for plan optimization and final dose verification. The accurate Monte Carlo (MC) dose algorithm is considered to be time-consuming for the optimization. Thus a fast, simplified dose algorithm is used in the optimization. The significant differences between the optimized dose and the delivered dose lead to tediously-planning loops and potentially-suboptimal solutions. This work aims to develop an IMRT optimization algorithm to minimize the dose discrepancy so that the delivered dose can be optimized in a holistic, end-to-end manner.Methods: The proposed algorithm, namely NeuralDAO, integrates a neural dose network into the Column-Generation (CG) Direct Aperture Optimization (DAO) formulation for step-and-shoot IMRT planning. The neural dose network is designed and trained to produce doses of MC-level accuracy within few milliseconds. Its differentiability is fully exploited to compute gradients for identifying potential aperture shapes. A prototype of NeuralDAO was developed in PyTorch and available to the public. Five lung patient cases have been studied. Dosimetric accuracy was compared with the MC dose. Plan quality and time were compared with a state-of-the-art (SoA) dose-correct algorithm. Statistical analysis was performed by Wilcoxon signed-rank test.Results: The average gamma passing rate at 2mm2% is 99.7% between the optimized and delivered dose. The Convergence process produced by NeuralDAO is virtually identical to that produced by an MC-based DAO. The average dose calculation time is 12.1 milliseconds for an aperture on GPU. One session of optimization took 10 minutes to 36 minutes. Compared with the SoA, a better conformity index and homogeneity index were observed for the target. The esophagus was significantly spared. Significant reductions were observed for the re-planning number and the planning time.Conclusions: A new DAO algorithm based on the neural dose network has been developed. The results suggest this algorithm minimizes the discrepancy between the optimized and delivered dose, which offers a promising approach to reduce the time and effort required in IMRT planning. This work demonstrates the possibility of applying the neural network in IMRT optimization. It is of great potential to extend this algorithm to other treatment modalities.This article is protected by copyright. All rights reserved
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?