Physical origin of the peak tailing of monoclonal antibodies in size-exclusion chromatography using bio-compatible systems and columns

Gritti, Fabrice
DOI: https://doi.org/10.1007/s00216-023-05119-2
2024-01-19
Analytical and Bioanalytical Chemistry
Abstract:The analysis of mixtures containing monoclonal antibody (mAb) (approximately 150 kDa molecular weight) and sub-unit impurities (approximately 100 kDa) is challenging, even when adopting the latest ultra-high-pressure liquid chromatography (UHPLC) columns (4.6 mm 150 mm coated hardware, 1.7 m 250 BEH Surface-modified Particles) and systems (ACQUITY UPLC I-class Bio Plus). The main issue still encountered is a persistent tail of the mAb peak. Here, the physical origin(s) of such peak tailing in size-exclusion chromatography (SEC) are investigated from both fundamental and practical approaches. Up to five relevant physical origins are analyzed: sample heterogeneity (glycoforms), UHPLC system dispersion, strong residual binding of the mAb to the SEC particles (via hydrophobic and/or electrostatic interactions) and to the stainless steel column/system hardware, slow escape kinetics of the mAb from the SEC particles, and flow heterogeneity caused by the non-ideal slurry packing of SEC columns. Experiments (testing sample heterogeneity, system dispersion, and strong residual interactions) and calculations (predicting the transient absorption/escape kinetics in a single SEC particle and the two-dimensional peak concentration profiles) altogether unambiguously demonstrate that the observed mAb peak tailing is caused primarily by the long-range velocity biases across the SEC column combined with the slow transverse dispersion of mAbs. Therefore, improvement in the resolution between mAb and sub-unit fragment impurities can only be achieved by increasing the column length, e.g., by applying recycling chromatography at acceptable pressures.
chemistry, analytical,biochemical research methods
What problem does this paper attempt to address?