A novel haemodilution chip mounted on the total thrombus‐formation analysis system, a flow chamber system, enables stable analysis of platelet products under low platelet concentration/high shear conditions

Akihiro Fuchizaki,Kazuta Yasui,Tomoya Hayashi,Mitsunobu Tanaka,Chiaki Oyamada,Tomoko Ohnishi‐Wada,Kazuya Hosokawa,Yoshihiro Fujimura,Kazushige Shimogaki,Fumiya Hirayama,Takafumi Kimura,Yoshihiro Takihara
DOI: https://doi.org/10.1111/vox.13683
2024-05-26
Vox Sanguinis
Abstract:Background and Objectives The total thrombus‐formation analysis system (T‐TAS) can quantitatively analyse the contribution of platelets to haemostasis using reconstituted blood samples. However, it is unsuitable in cases with low platelet counts. We introduced a haemodilution (HD) chip with a shallow chamber depth, adapted to low platelet counts and high shear conditions (1500 s−1). Materials and Methods Blood samples were prepared by mixing red blood cell products, standard human plasma and platelet products; the final platelet count was 50 × 103/μL. Aggregation tests were performed by using the aggregation inducers collagen, adenosine diphosphate (ADP) and ristocetin. Samples with 2‐, 4‐ and 9‐day‐old platelet products (N = 10) were evaluated. Results The HD chip enabled the stable analysis of the haemostatic function of all samples at a platelet count of 50 × 103/μL. Haemostatic function was correlated with ADP aggregation (time to 10 kPa [T10]: r = −0.53; area under the curve for 30 min: r = 0.40) and storage period (T10: r = 0.44). Conclusion The HD chip‐mounted T‐TAS can stably analyse haemostatic function under low platelet counts and high shear conditions; this approach is expected to serve as a bridge to in vivo haemostatic tests with experimental animals.
hematology
What problem does this paper attempt to address?