How much different is the semi-quantification of synovitis according to the ultrasound system and the blood flow detection technology?

Annamaria Iagnocco,Lina Martínez-Estupiñán,Fabiana Figus,Otto Olivas-Vergara,Alen Zabotti,Pablo Borges,Cecilia Agnes,Anna Zanetti,Davide Rozza,Esperanza Naredo
DOI: https://doi.org/10.11152/mu-4391
2024-06-11
Medical Ultrasonography
Abstract:Aim: To compare synovial blood flow scoring between different technologies and ultrasound (US) systems in active and inactive rheumatoid arthritis (RA).Material and methods. Fifty-nine RA patients underwent B-mode, power Doppler (PD), colour Doppler (CD), B-Flow and High-Resolution (High-Res) PDI assessments of 6 joints with two US systems at two European centres. Each joint was semi-quantitatively scored for all ultrasound parameters. PD, CD and High-Res PDI synovial signal was also quantitatively scored.Results. Correlations between the total score of SH with system 1 and 2 were very high (≥ 0.90, p<0.0001). Baseline correlations between systems for PD and CD total scores were moderate to very high (0.44-0.96, p 0,05). B-Flow and High-Res total scores were significantly lower than PD or CD total scores (p<0.05).Conclusion. A high-end and an entry-level US system were interchangeable for scoring SH and showed similar sensitivity and responsiveness in scoring synovial blood flow by PD and CD but not interchangeability. B-Flow and High-Res PDI were responsive, but they showed different sensitivity to detect synovial blood flow compared to conventional Doppler
radiology, nuclear medicine & medical imaging,acoustics