Development of an organic polymer monolith column for the nano liquid chromatography fast analysis of monoclonal antibody in infusion bags prepared in a hospital pharmacy

Claire Andre,Yves Claude Guillaume
DOI: https://doi.org/10.1002/bmc.5940
2024-06-24
Biomedical Chromatography
Abstract:Poly(butyl methacrylate‐co‐ethylene dimethacrylate) monolith was in situ prepared in a liquid chromatography capillary column with a 75 μm internal diameter. This monolith offered high permeability (5.3 ± 10−14 m2) and good peak capacity (140 for a 15 cm column length at 300 nl/min with a 20 min gradient time). This is exemplified by its separation ability in reversed mode for subunit analysis of monoclonal antibodies after IdeS digestion (middle‐up analysis). The potential of this column was also illustrated for the fast analytical control of therapeutic monoclonal antibodies in standardized infusion bags prepared in advance in a pharmacy department. Linearity analysis revealed the column's capability for accurate quantification analysis of the different dose bandings (in mg) of monoclonal antibodies in <2 min. In addition, lifetime analysis data indicated that the column can be highly reproducible and has a long lifetime with stable and low back pressure. The variations observed on the peak shape and area between unstressed (intact) and stressed monoclonal antibodies indicated that our nano liquid chromatographic method was stability indicating. In addition, using a gradient elution mode, the presence of minor components in the infusion bags was visualized.
pharmacology & pharmacy,biochemistry & molecular biology,biochemical research methods,chemistry, analytical