Highly sensitive photoelectrochemical immunosensor for detecting cancer marker CA19-9 based on a new SnSe quantum dot

Johari-Ahar, Mohammad
DOI: https://doi.org/10.1007/s00604-023-05718-x
2023-03-25
Mikrochimica Acta
Abstract:A sandwich-type photoelectrochemical (PEC) immunosensor was constructed on a screen-printed electrode (SPE) using gold-coated tin selenide quantum dots (Au-SnSe QDs) to determine the carbohydrate antigen 19 9 (CA19-9). Water-soluble Au-SnSe QDs were prepared by coating low-cost SnSe QDs, prepared by reacting tin(II) 2-ethyl hexanoate with selenium ions (HNaSe) without needing to add an external capping agent (SnSe QDs). SnSe-based QDs were characterized using high-resolution transmission electron microscopy (HR-TEM) and dynamic light scattering (DLS). DSP (dithio-bis (succinimidyl propionate)) as a linker was attached on Au@SnSe QDs and conjugated with CA19-9 monoclonal antibodies (Ab 2 -DSP-Au@SnSE QD). For capture probe assembling, an Au nano-layer was electrochemically deposited on a SPE by HAuCl 4 reduction using 12 cycles of cyclic voltammetry (0 to − 1.4 V) at the scan rate of 50 mV s −1 , then covered by self-assembly of DSP and covalent conjugation of CA19-9 Ab 1 . Our developed PEC immunosensor showed a significant photoelectrochemical response, recorded using chronoamperometry (0.3 V), for the presence of CA19-9 antigen in serum samples under light irradiation, with a detection limit (LOD) of 0.0011 U mL −1 and a dynamic range of 0.005–100 U mL −1 . The recovery of CA19-9 determination from serum samples was 101 to 113%.
What problem does this paper attempt to address?