Simulation of daily soft multifocal contact lenses using SimVis Gekko: from in-vitro and computational characterization to clinical validation

Eduardo Esteban-Ibañez,Diego Montagud-Martínez,Lucie Sawides,Amal Zaytouny,Alberto de Castro,Irene Sisó-Fuertes,Xoana Barcala,David P. Piñero,Walter D. Furlan,Carlos Dorronsoro,Enrique Gambra
DOI: https://doi.org/10.1038/s41598-024-59178-1
IF: 4.6
2024-04-15
Scientific Reports
Abstract:Multifocal contact lenses (MCLs) are one of the solutions to correct presbyopia, but their adoption is not widespread. To address this situation, visual simulators can be used to refine the adaptation process. This study aims to obtain accurate simulations for a visual simulator (SimVis Gekko; 2EyesVision) of daily soft MCL designs from four manufacturers. In-vitro characterization of these MCLs—several powers and additions- was obtained using NIMO TR-1504. From the averaged relative power profiles across powers, phase maps were reconstructed and the Through-Focus Visual Strehl metric was calculated for each MCL design. The SimVis Gekko simulation corresponding to each MCL design was obtained computationally and bench-validated. Finally, the MCL simulations were clinically validated involving presbyopic patients. The clinical validation results show a good agreement between the SimVis Gekko simulations and the real MCLs for through-focus visual acuity (TF-VA) curves and VA at three real distances. All MCL designs showed a partial correlation higher than 0.90 and a Root Mean Square Error below 0.07 logMAR between the TF-VA of simulations and Real MCLs across subjects. The validity of the simulation approach using SimVis Gekko and in-vitro measurements was confirmed in this study, opening the possibility to accelerate the adaptation of MCLs.
multidisciplinary sciences
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to improve the adaptability and clinical application effect of multifocal contact lenses (MCLs). Specifically, the main objective of the research is to accurately simulate daily soft multifocal contact lenses of different brands through a visual simulator (SimVis Gekko), and ensure the validity of these simulations through in - vitro and computational characterization as well as clinical verification. This helps ophthalmologists show patients the advantages and disadvantages of different MCL designs non - invasively, thereby improving the patient's adaptation process and increasing the patient's acceptance of MCLs. ### Research Background With the aging of the global population, the prevalence of presbyopia is on the rise continuously, and it is expected to reach 2.1 billion people by 2030. Presbyopia is a decrease in near vision caused by a decline in the elasticity of the lens, usually starting around the age of 45 and gradually worsening with age. Currently, methods for correcting presbyopia include glasses, intraocular lenses (IOLs), contact lenses (CLs), and corneal laser refractive surgery (PresbyLASIK), etc. Although multifocal contact lenses (MCLs) are available on the market, the proportion of presbyopic patients using MCLs is still low, mainly due to dry eyes, non - adaptation, poor visual performance, and a complicated fitting process. ### Research Objectives The main objective of this study is to obtain accurate SimVis Gekko simulations of daily soft multifocal contact lenses of different brands through the following steps: 1. **In - vitro and Computational Characterization**: Use the NIMO TR - 1504 device to measure the power distribution of MCLs, and calculate the phase map and the through - focus visual Strehl (TF - VS) index through a dedicated algorithm. 2. **Clinical Verification**: Conduct clinical verification among a small number of volunteers, and compare the through - focus visual acuity (TF - VA) curve of SimVis Gekko simulations with that of actual MCLs and the visual acuity (VA) at three actual distances. ### Methods 1. **In - vitro Characterization of MCLs**: Use the NIMO TR - 1504 device to measure the power distribution of four different brands of MCLs (MyDay, 1 - Day Acuvue Moist, Dailies Total1, and Biotrue ONEday). 2. **Computational Process**: Calculate the phase map and TF - VS based on the average power distribution for generating SimVis Gekko simulations. 3. **Verification of SimVis Gekko Simulations**: Verify the accuracy of SimVis Gekko simulations through high - speed lensometers experiments. 4. **Clinical Verification**: Conduct clinical verification among 8 presbyopic patients, and compare the visual performance of SimVis Gekko simulations with that of actual MCLs. ### Results 1. **In - vitro Characterization Results**: 42 power distributions were measured, and the repeatability measurement standard deviation of each MCL design was less than 0.10 D. MCLs of different brands showed different design characteristics, such as center - near design and concentric - ring design. 2. **Computational Results**: The TF - VS obtained through calculation shows that as the optical diameter increases, the depth of field (DoF) increases, and the main peak shifts towards the hyperopia direction. 3. **Clinical Verification Results**: The TF - VA curve and VA of SimVis Gekko simulations and actual MCLs are highly consistent in performance at three actual distances, with the partial correlation coefficient (rxy,z) greater than 0.90 and the root - mean - square error (RMSE) lower than 0.07 logMAR. ### Discussion This study successfully obtained accurate SimVis Gekko simulations of daily soft multifocal contact lenses of different brands through in - vitro and computational characterization, and confirmed the validity of these simulations through clinical verification. This provides ophthalmologists with a powerful tool to show patients the advantages and disadvantages of different MCL designs non - invasively, thereby improving the patient's adaptation process and increasing the acceptance of MCLs.