Viewpoint Planning for Quantitative Coronary Angiography

Alexander Preuhs,Martin Berger,Sebastian Bauer,Thomas Redel,Mathias Unberath,Stephan Achenbach,Andreas Maier
DOI: https://doi.org/10.48550/arXiv.1808.00853
IF: 4.506
2018-08-02
Medical Physics
Abstract:In coronary angiography the condition of myocardial blood supply is assessed by analyzing 2-D X-ray projections of contrasted coronary arteries. This is done using a flexible C-arm system. Due to the X-ray immanent dimensionality reduction projecting the 3-D scene onto a 2-D image, the viewpoint is critical to guarantee an appropriate view onto the affected artery and, thus, enable reliable diagnosis. In this work we introduce an algorithm computing optimal viewpoints for the assessment of coronary arteries without the need for 3-D models. We introduce the concept of optimal viewpoint planning solely based on a single angiographic X-ray image. The subsequent viewpoint is computed such that it is rotated precisely around a vessel, while minimizing foreshortening. Our algorithm reduces foreshortening substantially compared to the input view and completely eliminates it for 90 degree rotations. Rotations around iso-centered foreshortening-free vessels passing the isocenter are exact. The precision, however, decreases when the vessel is off-centered or foreshortened. We evaluate worst case boundaries, providing insight in the maximal inaccuracies to be expected. This can be utilized to design viewpoints guaranteeing desired requirements, e.g. a true rotation around the vessel of at minimum 30 degree.
What problem does this paper attempt to address?