A multisite validation of a two hours antibiotic susceptibility flow cytometry assay directly from positive blood cultures

C. Pina-Vaz,A. Silva-Dias,I. Martins-Oliveira,R. Gomes,B. Perez-Viso,S. Cruz,A. G. Rodrigues,A. Sarmento,R. Cantón
DOI: https://doi.org/10.1186/s12866-024-03341-1
IF: 4.465
2024-05-29
BMC Microbiology
Abstract:Rapid antimicrobial susceptibility testing (AST) is urgently needed to provide safer treatment to counteract antimicrobial resistance. This is critical in septic patients, because resistance increases empiric therapy uncertainty and the risk of a poor outcome. We validate a novel 2h flow cytometry AST assay directly from positive blood cultures (PBC) by using a room temperature stable FASTgramneg and FASTgrampos kits (FASTinov® Porto, Portugal) in three sites: FASTinov (site-1), Hospital Ramon y Cajal, Madrid, Spain (site-2) and Centro Hospitalar S. João, Porto, Portugal (site-3). A total of 670 PBC were included: 333 spiked (site-1) and 337 clinical PBC (151 site-2 and 186 site-3): 367 gram-negative and 303 gram-positive. Manufacturer instructions were followed for sample preparation, panel inoculation, incubation (1h/37oC) and flow cytometry analysis using CytoFlex (Site-1 and -2) or DxFlex (site-3) both instruments from Beckman-Coulter, USA.