Quantitative evaluation of the pivot shift by image analysis using the iPad

Yuichi Hoshino, Paulo Araujo, Mattias Ahldén, Kristian Samuelsson, Bart Muller, Marcus Hofbauer, Megan R Wolf, James J Irrgang, Freddie H Fu, Volker Musahl
2013-04-01
Abstract:Purpose To enable comparison of test results, a widely available measurement system for the pivot shift test is needed. Simple image analysis of lateral knee joint translation is one such system that can be installed on a prevalent computer tablet (e.g. iPad). The purpose of this study was to test a novel iPad application to detect the pivot shift. It was hypothesized that the abnormal lateral translation in ACL deficient knees would be detected by the iPad application. Methods Thirty-four consecutive ACL deficient patients were tested. Three skin markers were attached on the following bony landmarks: (1) Gerdy’s tubercle, (2) fibular head and (3) lateral epicondyle. A standardized pivot shift test was performed under anaesthesia, while the lateral side of the knee joint was monitored. The recorded movie was processed by the iPad application to measure the …