First ground-level laboratory test of the two-phase xenon emission detector RED-100

D.Y. Akimov,V.A. Belov,A.I. Bolozdynya,A.G. Dolgolenko,Y.V. Efremenko,A.V. Etenko,A.V. Galavanov,D.V. Gouss,Y.V. Gusakov,D.E. Kdib,A.V. Khromov,A.M. Konovalov,V.N. Kornoukhov,A.G. Kovalenko,E.S. Kozlova,A.V. Kumpan,A.V. Lukyashin,Y.A. Melikyan,V.V. Moramzin,D.A. Nischeta,O.E. Razuvaeva,D.G. Rudik,A.V. Shakirov,G.E. Simakov,V.V. Sosnovtsev,Y.V. Stogov,A.A. Vasin
DOI: https://doi.org/10.1088/1748-0221/15/02/P02020
2020-02-21
Journal of Instrumentation
Abstract:RED-100 is a two-phase detector for study of coherent elastic scattering of reactor electron antineutrinos off xenon atomic nuclei. The detector contains a total of 200 kg of liquid xenon in a titanium cryostat with 160 kg of xenon in active volume inside a Teflon-made light collection cage associated with electrode system. The active volume is viewed by two arrays of nineteen 3''-diameter Hamamatsu R11410-20 PMTs assembled in two planes on top and bottom. The electrode system is equipped with an electron shutter (a patented device) to reduce a "spontaneous" single-electron noise. The detector was tested in a ground-level laboratory. The obtained results demonstrate that detection of coherent elastic scattering of reactor antineutrinos off xenon nuclei at Kalinin nuclear power plant with the RED-100 detector is feasible with a threshold of 4 ionization electrons.
instruments & instrumentation
What problem does this paper attempt to address?