Implant Fracture Analysis of the TFNA Proximal Femoral Nail

Anton Lambers,Bertram Rieger,Alan Kop,Peter D'Alessandro,Piers Yates
DOI: https://doi.org/10.2106/JBJS.18.00997
2019-10-07
Abstract:Background: Mechanical failure of intramedullary nails is rare. This investigation was prompted by a series of cases of observed breakage of the recently introduced TFNA Proximal Femoral Nailing System (TFN [Trochanteric Fixation Nail]-Advanced; DePuy Synthes) in our region. Laboratory analysis and case data are presented, in contribution to post-market surveillance. Methods: Medical and imaging records from the 3 public tertiary orthopaedic trauma hospitals in Western Australia were reviewed. Relevant data of patients in whom breakage of the TFNA implant occurred between 2016 and 2018 were collected and analyzed. Laboratory review of retrieved implants was conducted at the Centre for Implant Technology and Retrieval Analysis (CITRA) in Western Australia.