Design and optimization of a breast-dedicated SPECT scanner with multi-lofthole collimation

M. Saed,A. Sadremomtaz and H. Mahani
DOI: https://doi.org/10.1088/1748-0221/17/01/P01006
2022-01-04
Journal of Instrumentation
Abstract:Background : the need for simultaneous high-sensitivity and high-resolution breast SPECT imaging mandates to design and optimize dedicated scanners. Therefore, this work aims to design and optimize a novel breast-dedicated SPECT system with multi-lofthole collimator. Materials and Methods : in this research, a novel breast-dedicated scanner is designed and then optimized. The scanner is equipped with a single full-ring multi-lofthole collimation long with modular NaI(Tl) detectors. The step-and-shoot data acquisition was considered with two steps. Then, an analytic optimization was conducted to balance the existing sensitivity-resolution tradeoff. To do so, several scanner geometries were investigated. The optimal configuration maximized the system sensitivity at a given system resolution. Furthermore, the scanner was also modeled within the GATE simulator. Then, detector energy resolution, septal penetration and scattering, and system sensitivity were calcu...
instruments & instrumentation
What problem does this paper attempt to address?