Utility and usability of laser speckle contrast imaging (LSCI) for displaying real-time tissue perfusion/blood flow in robot-assisted surgery (RAS): comparison to indocyanine green (ICG) and use in laparoscopic surgery

Yao Z. Liu,Shinil K. Shah,Christina M. Sanders,Chibueze A. Nwaiwu,Alyson F. Dechert,Saloni Mehrotra,Steven D. Schwaitzberg,Peter C. W. Kim,Erik B. Wilson
DOI: https://doi.org/10.1007/s00464-022-09590-3
2022-09-16
Surgical Endoscopy
Abstract:Utility and usability of laser speckle contrast imaging (LSCI) in detecting real-time tissue perfusion in robot-assisted surgery (RAS) and laparoscopic surgery are not known. LSCI displays a color heatmap of real-time tissue blood flow by capturing the interference of coherent laser light on red blood cells. LSCI has advantages in perfusion visualization over indocyanine green imaging (ICG) including repeat use on demand, no need for dye, and no latency between injection and display. Herein, we report the first-in-human clinical comparison of a novel device combining proprietary LSCI processing and ICG for real-time perfusion assessment during RAS and laparoscopic surgeries.
What problem does this paper attempt to address?