Peptide-anchored biomimetic interface for electrochemical detection of cardiomyocyte-derived extracellular vesicles

Yang Zhou,Fei Zhao,Bo Zheng,Shihai Tang,Juan Gong,Bin He,Zhi Zhang,Na Jiang,Huijuan Zha,Jun Luo
DOI: https://doi.org/10.1007/s00216-022-04419-3
2022-11-13
Analytical and Bioanalytical Chemistry
Abstract:Cardiomyocyte-derived extracellular vesicles (EVs) are a promising class of biomarkers that can advance the diagnosis of many kinds of cardiovascular diseases. Herein, we develop a new electrochemical method for the feasible detection of cardiomyocyte-derived EVs in biological fluids. The core design of the method is the fabrication of a peptide-anchored biomimetic interface consisting of a lipid bilayer and peptide probes. On the one hand, the lipid bilayer provides excellent antifouling ability to the electrode interface and facilitates the anchoring of peptide probes. On the other hand, the peptide probes equip the electrode interface with excellent binding capability and affinity to CD172a, a specific marker of cardiomyocyte-derived EVs, thus enabling the efficient and selective detection of target EVs. Taking EVs derived from the heart myoblast cells H9C2 as the model target, the method displays a wide linear detection range from 1 × 10 3 to 1 × 10 8 particles/mL with a desirable detection limit of 132 particles/mL. Furthermore, the method shows good performance in biological fluids such as serum, and thus may have great potential for practical use in the diagnosis of cardiovascular diseases.
What problem does this paper attempt to address?