Sensitivity analysis of magnetic field measurements for magnetic resonance electrical impedance tomography (MREIT)
Cihan Göksu,Klaus Scheffler,Philipp Ehses,Lars G. Hanson,Axel Thielscher
DOI: https://doi.org/10.1002/mrm.26727
2017-05-30
Magnetic Resonance in Medicine
Abstract:PURPOSE: Clinical use of magnetic resonance electrical impedance tomography (MREIT) still requires significant sensitivity improvements. Here, the measurement of the current-induced magnetic field (ΔB<sub>z,c</sub> ) is improved using systematic efficiency analyses and optimization of multi-echo spin echo (MESE) and steady-state free precession free induction decay (SSFP-FID) sequences.THEORY AND METHODS: Considering T<sub>1</sub> , T<sub>2</sub> , and T2* relaxation in the signal-to-noise ratios (SNRs) of the MR magnitude images, the efficiency of MESE and SSFP-FID MREIT experiments, and its dependence on the sequence parameters, are analytically analyzed and simulated. The theoretical results are experimentally validated in a saline-filled homogenous spherical phantom with relaxation parameters similar to brain tissue. Measurement of ΔB<sub>z,c</sub> is also performed in a cylindrical phantom with saline and chicken meat.RESULTS: The efficiency simulations and experimental results are in good agreement. When using optimal parameters, ΔB<sub>z,c</sub> can be reliably measured in the phantom even at injected current strengths of 1 mA or lower for both sequence types. The importance of using proper crusher gradient selection on the phase evolution in a MESE experiment is also demonstrated.CONCLUSION: The efficiencies observed with the optimized sequence parameters will likely render in-vivo human brain MREIT feasible. Magn Reson Med 79:748-760, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
radiology, nuclear medicine & medical imaging