Characterization and Response Model of the PPS-M Aerosol Sensor

Antti Rostedt,Anssi Arffman,Kauko Janka,Jaakko Yli-Ojanperä,Jorma Keskinen
DOI: https://doi.org/10.1080/02786826.2014.951023
2014-08-14
Aerosol Science and Technology
Abstract:The Pegasor PPS-M sensor is an electrical aerosol sensor based on diffusion charging and current measurement without particle collection. In this study, the role and effect of each component in the instrument is discussed shortly and the results from a thorough calibration measurements are presented. A comprehensive response model for the operation of the PPS-M sensor was developed based on the calibration results and computational fluid dynamics (CFD) modeling results. The obtained response model, covering the effects of the particle charger, the mobility analyzer, and both diffusion and inertial losses, was tested in the laboratory measurements with polydisperse test aerosols, where a good correlation between the model and the measured results was found. Copyright 2014 American Association for Aerosol Research
environmental sciences,meteorology & atmospheric sciences,engineering, chemical, mechanical
What problem does this paper attempt to address?