Correction of the effects of carbon dioxide and hydrogen sulfide on electrostatic cell monitors measurements of radon in water

Robert-Csaba Begy,Codrin-Fabian Savin,Alida Timar-Gabor
DOI: https://doi.org/10.1016/j.jece.2021.107040
IF: 7.7
2022-02-01
Journal of Environmental Chemical Engineering
Abstract:The present study shows that the presence of hydrogen sulfide and carbon dioxide in water have a negative effect on the accuracy of radon measurements in samples when an electrostatic ion collector device (RAD7) is used.CO2 concentrations between 0.65 g/l and 5.19 g/l were generated in water samples containing 222Rn activities ranging from 23.4 ± 0.9 Bq/l to 6.7 ± 0.9 Bq/l, after which the samples were measured by a RAD7 device, and the results compared to the radon activity values of the gas-free samples. The same procedure was followed for H2S concentrations from 0.125 mg/l to 3 mg/l. In both cases, a major impact was observed, the radon apparent specific activities decreasing exponentially with gas concentration. For a CO2 concentration of 1.29 g/l, the apparent specific activity underestimated the true value by 24.5%, while for a concentration of 5.19 g/l CO2 the decrease was of 57%. For H2S, the radon measured values decreased by 42.5% for concentrations as low as 0.5 mg/l, culminating in a reduction of 86.9% for 3 mg/l H2S. We are attributing these effects to 218Po neutralization processes, which reduces the electrostatic efficiency of the silicon detector system. To eliminate the interferences of these gases on RAD7 measurements, a sample preparation method was developed and tested, involving the precipitation and deposition of CO2 using Ca(OH)2, and H2S using Pb(NO3)2..The findings of the present study clearly indicate that in order to perform valid radon measurements in gas-rich waters with RAD7, the effect of the interfering gases must be considered.
engineering, chemical, environmental
What problem does this paper attempt to address?