Impact of image denoising on image quality, quantitative parameters and sensitivity of ultra-low-dose volume perfusion CT imaging

Ahmed E. Othman,Carolin Brockmann,Zepa Yang,Changwon Kim,Saif Afat,Rastislav Pjontek,Omid Nikoubashman,Marc A. Brockmann,Konstantin Nikolaou,Martin Wiesmann,Jong Hyo Kim
DOI: https://doi.org/10.1007/s00330-015-3853-6
IF: 7.034
2015-05-30
European Radiology
Abstract:ObjectivesTo examine the impact of denoising on ultra-low-dose volume perfusion CT (ULD-VPCT) imaging in acute stroke.MethodsSimulated ULD-VPCT data sets at 20 % dose rate were generated from perfusion data sets of 20 patients with suspected ischemic stroke acquired at 80 kVp/180 mAs. Four data sets were generated from each ULD-VPCT data set: not-denoised (ND); denoised using spatiotemporal filter (D1); denoised using quanta-stream diffusion technique (D2); combination of both methods (D1 + D2). Signal-to-noise ratio (SNR) was measured in the resulting 100 data sets. Image quality, presence/absence of ischemic lesions, CBV and CBF scores according to a modified ASPECTS score were assessed by two blinded readers.ResultsSNR and qualitative scores were highest for D1 + D2 and lowest for ND (all p ≤ 0.001). In 25 % of the patients, ND maps were not assessable and therefore excluded from further analyses. Compared to original data sets, in D2 and D1 + D2, readers correctly identified all patients with ischemic lesions (sensitivity 1.0, kappa 1.0). Lesion size was most accurately estimated for D1 + D2 with a sensitivity of 1.0 (CBV) and 0.94 (CBF) and an inter-rater agreement of 1.0 and 0.92, respectively.ConclusionAn appropriate combination of denoising techniques applied in ULD-VPCT produces diagnostically sufficient perfusion maps at substantially reduced dose rates as low as 20 % of the normal scan.Key Points• Perfusion-CT is an accurate tool for the detection of brain ischemias.• The high associated radiation doses are a major drawback of brain perfusion CT.• Decreasing tube current in perfusion CT increases image noise and deteriorates image quality.• Combination of different image-denoising techniques produces sufficient image quality from ultra-low-dose perfusion CT.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?