A 7T spine array based on electric dipole transmitters

Qi Duan,Govind Nair,Natalia Gudino,Jacco A de Zwart,Peter van Gelderen,Joe Murphy-Boesch,Daniel S Reich,Jeff H Duyn,Hellmut Merkle,Jacco A. de Zwart,Daniel S. Reich,Jeff H. Duyn
DOI: https://doi.org/10.1002/mrm.25817
2015-07-20
Magnetic Resonance in Medicine
Abstract:PURPOSE: The goal of this study was to explore the feasibility of using an array of electric dipole antennas for RF transmission in spine MRI at high fields.METHOD: A two-channel transmit array based on an electric dipole design was quantitatively optimized for 7T spine imaging and integrated with a receive array combining eight loop coils. Using B1+ mapping, the transmit efficiency of the dipole array was compared with a design using quadrature loop pairs. The radiofrequency energy deposition for each array was measured using a home-built dielectric phantom and MR thermometry. The performance of the proposed array was qualitatively demonstrated in human studies.RESULTS: The results indicate dramatically improved transmit efficiency for the dipole design compared with the loop excitation. A gain of up to 76% was achieved within the spinal region.CONCLUSION: For imaging of the spine, electric dipole-based transmitters provide an attractive alternative to the traditional loop-based design. Easy integration with existing receive array technology facilitates practical use at high fields.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?