Practical approach to blood compatibility assessments: general considerations and standards

M.F. Wolf,J.M. Andwraon
DOI: https://doi.org/10.1533/9780857096456.2.159
2012-01-01
Abstract:Medical devices and medical device materials must be designed to achieve specific criteria for safety and performance. The criteria themselves may be bound by physical, chemical, and material limitations, and by biochemical, molecular, and cellular processes within the body recognized to take place upon contact. In the latter case, responses can be favorable or unfavorable, partially controlled by learned approaches, for example drugs, or be uncontrollable by current experience and know-how. As such, determining what testing is required and which test results are acceptable and meaningful in the development of a new medical device can be complex. This chapter highlights some of the main considerations in the specific area of evaluation of blood-contacting medical devices and materials. It discusses the main factors that may impact the blood-device/ material responses that can occur, and it provides examples of some of the most common observations and approaches to manage the response. As the international standard ISO 10993–4 has been and continues to be a major driving influence behind this aspect of the biological safety evaluation of materials and devices, the evolving direction of the standard is covered and controversies in testing are reviewed. Ultimately, appropriate care taken to evaluate and understand critical blood-device/material interactions in each unique application leads to predictive and safe use in humans.