Multicentre evaluation of the BYG Carba v2.0 test, a simplified electrochemical assay for the rapid laboratory detection of carbapenemase-producing Enterobacteriaceae

Pierre Bogaerts,Saoussen Oueslati,Danièle Meunier,Claire Nonhoff,Sami Yunus,Marion Massart,Olivier Denis,Neil Woodford,Katie L. Hopkins,Thierry Naas,Laurent Dortet,Te-Din Huang,Youri Glupczynski
DOI: https://doi.org/10.1038/s41598-017-09820-y
IF: 4.6
2017-08-30
Scientific Reports
Abstract:Rapid detection of carbapenemase-producing Enterobacteriaceae (CPE) represents a major challenge for microbiology laboratories. We evaluated the BYG Carba v2.0 using a simplified protocol, which detects CPE in less than 30 minutes. This new procedure reduces the hands-on-time from 5 to one minute and only requires a limited amount of material (one to three colonies) thereby preventing the need for subculturing bacterial isolates to reach a larger amount of pure biomass. This multicentre study involved four European reference laboratories. For the 1181 isolates tested across four centres, BYG Carba v2.0 yielded overall sensitivity and specificity of 96.3% (CI95: 94.5–97.5) and 99.7% (CI95: 98.6–100) respectively. Considering only the 670 consecutive isolates tested prospectively, the BYG Carba v2.0 displayed overall positive and negative predictive values of 99.7% (CI95: 95.4–98.9) and 97.5% (CI95: 94.9–98.8). Regarding time to positivity, 85% of CPE detected were positive within ten minutes. The BYG Carba v2.0 is a new highly simplified, rapid and accurate electrochemical assay discriminating between CPE and non-CPE in less than 30 min. The real-time quantified signal allows objective and traceable interpretation of the results.
multidisciplinary sciences
What problem does this paper attempt to address?