Field dependence of electronic recoil signals in a dual-phase liquid xenon time projection chamber

E. Hogenbirk,M.P. Decowski,K. McEwan,A.P. Colijn
DOI: https://doi.org/10.1088/1748-0221/13/10/p10031
2018-10-26
Journal of Instrumentation
Abstract:We present measurements of light and charge signals in a dual-phase time projection chamber at electric fields varying from 10 V/cm up to 500 V/cm and at zero field using 511 keV gamma rays from a 22Na source. We determine the drift velocity, electron lifetime, diffusion constant, and light and charge yields at 511 keV as a function of the electric field. In addition, we fit the scintillation pulse shape to an effective exponential model, showing a decay time of 43.5 ns at low field that decreases to 25 ns at high fields.