Waste Management Control by Applying Novel Method for HPLC Technique to Replace SKALAR Cadmium Reduction Method for Nitrate and Nitrite in Water Analysis

Yahya AlOudah
DOI: https://doi.org/10.2523/iptc-22291-ea
2022-02-21
Abstract:Abstract A new method was developed on High Performance Liquid Chromatography (HPLC) instrument equipped with a Photodiode Array UV detector to determine the Nitrite-Nitrogen and Nitrate-Nitrogen in seawater, industrial wastewater, and groundwater at very low concentrations. This HPLC used a reverse phase system with polar mobile phase and nonpolar C-18 Column. The split mechanism of the HPLC Chromatography was accomplished by employing acetonitrile and acidified water with pH 2.5, at 55:45 ACN: H2O v/v as a mobile phase, long HPLC Column (C-18, 0.5 μm, 250 × 4.6 mm) and at 1 mL/min flow rate. The calibration curves of nitrate-nitrogen and nitrite-nitrogen in the range of 0.05 to 1.0 mg/L were exceptionally linear with correlation coefficients greater than 0.999 and with low Method Detection Limit (MDL) of 0.02 mg/L for both of NO2-n and NO3-n. This developed method also showed comparable results to the ones obtained by the traditional SKALAR instrument which follows ASTM D3867 Automated Cadmium Reduction Method. The results of the nitrate-n and nitrite-n for the Quality Control Samples (QCS) and the average recoveries percentage of the spiked samples demonstrated high accuracy values 99.20% - 101.40% & 98.4% - 99.1% respectively. The very low detection limit is what set this study apart from the previous conducted studies on nitrate and nitrite determination by HPLC. This new developed method is safer and with minimal cost compared to the traditional cadmium reduction method. It doesn't consume water, or require preparation of many hazardous reagents that contain and generate cadmium wastes.
What problem does this paper attempt to address?