Single- and dual-energy CT pulmonary angiography using second- and third-generation dual-source CT systems: comparison of radiation dose and image quality

Lukas Lenga,Franziska Trapp,Moritz H. Albrecht,Julian L. Wichmann,Addison A. Johnson,Ibrahim Yel,Tommaso D’Angelo,Christian Booz,Thomas J. Vogl,Simon S. Martin
DOI: https://doi.org/10.1007/s00330-018-5982-1
IF: 7.034
2019-01-21
European Radiology
Abstract:ObjectivesTo evaluate radiation exposure and image quality in matched patient cohorts for CT pulmonary angiography (CTPA) acquired in single- and dual-energy mode using second- and third-generation dual-source CT (DSCT) systems.MethodsWe retrospectively included 200 patients (mean age, 65.5 years ± 15.7 years) with suspected pulmonary embolism—equally divided into four study groups (n = 50) and matched by gender and body mass index. CTPA was performed with vendor-predefined second-generation (group A, 100-kV single-energy computed tomography (SECT); group B, 80/Sn140-kV dual-energy computed tomography (DECT)) or third-generation DSCT (group C, 100-kV SECT; group D, 90/Sn150-kV DECT) devices. Radiation metrics were assessed using a normalized scan range of 27.5 cm. For objective image quality evaluation, dose-independent figure-of-merit (FOM) contrast-to-noise ratios (CNRs) were calculated. Subjective image analysis included ratings for overall image quality, reader confidence, and image artifacts using five-point Likert scales.ResultsCalculations of the effective dose (ED) of radiation for a normalized scan range of 27.5 cm showed nonsignificant differences between SECT and DECT acquisitions for each scanner generation (p ≥ 0.253). The mean effective radiation dose was lower for third-generation groups C (1.5 mSv ± 0.8 mSv) and D (1.4 mSv ± 0.7 mSv) compared to second-generation groups A (2.5 mSv ± 0.9 mSv) and B (2.3 mSv ± 0.6 mSv) (both p ≤ 0.013). FOM-CNR measurements were highest for group D. Qualitative image parameters of overall image quality, reader confidence, and image artifacts showed nonsignificant differences among the four groups (p ≥ 0.162).ConclusionsThird-generation DSCT systems show lower radiation dose parameters for CTPA compared to second-generation DSCT. DECT can be performed with both scanner generations without radiation dose penalty or detrimental effects on image quality compared to SECT.Key Points• Radiation exposure showed nonsignificant differences between SECT and DECT for both DSCT scanner devices.• Dual-energy CTPA provides equivalent image quality compared to standard image acquisition.• Subjective image quality assessment was similar among the four study groups.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?